Sono-Trak Transit Time Ultrasonic Flow Meter

Sono-Trak™ Transit Time Ultrasonic
»Features »Specifications »Installation »Model Codes
 
    
»View Demo
»Documentation
Transit-Time
»Technology

EMCO's Sono-Trak™ Transit Time ultrasonic flow meter combines advanced non-invasive flow measurement technology with a versatile transmitter design to provide an accurate, maintenance-free, easy-to-install, and easy-to-use measuring system. The Sono-Trak™ Transit Time flow meter, using only one set of transducers, measures liquids for pipe sizes from 2 to 100".


Reliable and Rugged
  • High accuracy: ±0.5% of rate (factory tested); ±2.0% of rate (typical field installation); ±0.1% of rate repeatability
  • Time domain expansion technology improves signal processing and dramatically increases resolution
  • With no moving parts to wear over time, flow meter maintenance is non-existent
  • Rugged, NEMA 6-rated transducers are fully submersible
  • No pressure drop
  • FM and CSA approvals (pending)
Fast & Easy Installation
  • Typical installations take less than 30 minutes with the non-invasive, clamp-on design using all stainless steel quick clamps
  • No worries of flange requirements, pressure ratings, process liquid interaction issues, pipe vibrations, or upstream transitions
  • Heavy-duty, urethane-coated cables are pre-terminated to facilitate a speedy installation
  • Easy-to-read, blue backlit display allows for quick flowmeter set-up, featuring EMCO's patented menu-driven EZ-Logic™ user interface
Features
Flexible
  • The same Sono-Trak flow meter can be installed on line sizes from 2 to 100" (50 to 2540 mm)
  • Plug-and-play input/output module options enable custom electronics configuration and/or field retrofits/upgrades; choose two from: pulse, analog, relay, RS-232, RS-485, BTU, and data logger
  • Standard, built-in infrared capability for use with optional Windows® software

The compact, NEMA 4X-rated, wall-mounted transmitter has a large, easy-to-read backlit display, which alternates between flow rate and totalized flow with corresponding, user-selectable engineering units. The four-button keypad allows for quick flow meter set-up, featuring EMCO's patented menu-driven EZ-Logic™ user interface. Additionally, the standard infrared communication port can be used with the optional Windows® compatible software package and serial interface device to access all configuration and data recording features.

With buffet-style input/output (I/O) module options, the electronics configuration is easily customized. Two I/O sockets will accept any of seven offered modules: pulse, analog, relay, RS-232, RS-485, BTU, and data logger. The modules are "plug-and-play," enabling field upgrades and/or retrofits. The RS-485 module allows single-point viewing of multiple meters in a master/slave configuration. The BTU module accepts two RTD inputs and performs energy calculations. The data logger collects up to 100,000 data points in non-volatile memory and can be removed from the transmitter under power and connected to a standard PC serial port for downloading and data analysis. Full diagnostics are accessible through the display, the standard infrared port, and the optional serial communication modules.



Back to top



Specifications

Performance Specifications
Accuracy (linear ranges) Factory tested¹ ± 0.5% of flow rate
Typical field installation² ± 1 to 2% of flow rate
Repeatability ± 0.1% of flow rate
Response time Adjustable from .3 to 30 sec
¹Factory test conditions: water at 70°F, flow velocity > 1 ft/sec on a 6" stainless steel pipe

²Field performance is dependent on the accuracy of the pipe's internal diameter and wall thickness measurements, as well as straight run conditions

Back to top


Operating Specifications
Applied pipeline sizes 2 to 100" (50 to 2540 mm)
Measurable fluids Most common liquids from waste water to petrochemical products
Measurable flow velocities Liquid -40 to 40 ft/sec (-12 to 12 m/sec)
Process pressure Not applicable
Ambient temperature limit -40 to 140°F (-40 to 60°C)
Process temperature limit Standard 40 to 250°F (-40 to 121°C)
High temp option 40 to 400°F (-40 to 205°C)
Ambient humidity limit Up to 95% RH, non-condensing
Power requirements Field configurable for AC or DC 12-24 VDC

115 VAC, 50/60 Hz (± 10%)

230 VAC, 50/60 Hz (± 10%)
Power consumption Less than 5 W
Input/Output modules 4 to 20 mA output Field configurable as a 2-wire active or passive transmitter, 800 Ω maximum resistance
Frequency output Field configurable for 0 to 1,000 Hz or 0 to 10,000 Hz output, 20 mA max (50% duty cycle)
Dual relay output 175 V, ¼ A switch, 1 A carry current, 0.2 Ω resistance
RS485 interface Master/slave configuration, supports 57.6 kbd communications up to 1000'
RS232 interface Supports up to 57.6 kbd communication (19.2 kbd @ 50')
Dual RTD input Accepts two 1000 Ω RTD inputs for BTU measurement
Data logger Non-volatile storage of 100,000 time/date stamped data points, selectable from 1 to 1,000 seconds. Connects with PC serial port for data retrieval



Back to top


Physical Specifications
Materials Wetted parts None
Transducer housing 250°F Delrin® / Ultem® / Stainless steel; NEMA 6 (IP68)
Transducer housing 400°F Toron® /Vespel® / Stainless steel; NEMA 6 (IP68)
Sensor cables Polyurethane armored coaxial
Clamps Stainless Steel
Cable lengths Standard 25' (7.6 m)
Options 50' (15.2 m)

75' (22.8m)
Custom 76' to 1000' (23.1 m - 304.8 m)
Maximum transducer cable length 1000' (304.8 m)
Cable connection Standard submersible
Mounting Transducer External, adjustable clamp
Electronics Remote wall
Approvals CSA Class I Div II Groups A,B,C,D

Class II and Class III Groups E, F



Back to top



Installation


Installation Specifications
Straight run piping Condition Upstream Downstream
Typical 10 5
Minimum 5 3
Upstream valve and other piping configurations Consult EMCO rep or the Factory Consult EMCO rep or the Factory
Transducer mounting Vertical The recommended flow direction a vertical pipe is upwards.
Horizontal Transducer should be mounted so the pipe is full of liquid with no bubbles present.
Zero-point adjustment For best accuracy, a zero-point setting must be made with a full pipe at no flow during initial start-up.
Air/Gas entrapment Excess gas entrapment can scatter or reflect the sound wave, causing inaccurate measurement. Allowable air/gas entrapment is 2%.
Pipe material Good results Caution needed Do not use on
Most metallic Cast iron Cement
Plastic Lined pipes Pipes with internal deposits or severe corrosion
Sonic wave forms Line size Mounting options
2 to 2.5" (DN50 to DN65) "W"
3 to 12" (DN80 to DN300) "V"
14 to 100" (DN350 to DN2500 "Z"



Back to top



Model Codes


Category Description Suffix codes
Electronics 2-line backlit display with 4-button keypad ST-30 - - - - - -
I/O port 1 None - N - - - - -
4 to 20 mA output - A - - - - -
Frequency/pulse output - F - - - - -
Dual relay output (dry contact) - R - - - - -
Dual RTD (limit 1 per unit)* - BTU - - - - -
RS-232 interface ** - RS2 - - - - -
RS-485 interface ** - RS4 - - - - -
I/O port 2 None - - N - - - -
4 to 20 mA output - - A - - - -
Frequency/pulse output - - F - - - -
Dual relay output (dry contact) - - R - - - -
Dual RTD (limit 1 per unit)* - - BTU - - - -
RS-232 interface ** - - RS2 - - - -
RS-485 interface ** - - RS4 - - - -
Process temperature -40 to 250°F (-40 to 121°C) - - - 250 - - -
-40 to 400°F (-40 to 205°C) - - - 400 - - -
Transducers Submersible - - - - 1 - -
Cable length 25' (7.6 m) - - - - - 25 -
50' (15.2 m) - - - - - 50 -
75' (22.8 m) - - - - - 75 -
76 to 1000' (23.1 to 304.8 m) (specify length) - - - - - XX -
Pipe clamp size 2 to 12" (50 to 300 mm) - - - - - - 12
14 to 24" (350 to 600 mm) - - - - - - 24
26 to 36" (650 to 900 mm) - - - - - - 36
38 to 100" (950 to 2540 mm) (specify pipe size) - - - - - - XX